Description
- Product Detail
- Send Inquiry
The HP-HEMAD2000 haemodialysis machine is specially designed, produced and sold for renal failure patients to receive haemodialysis.It is applicable for chronic renal failure and other blood Purification treatment.
Choosing Reasons:1,Classical and perfect model designed. 2,Two persons can be trained freely for first time. 4,All accessories or spare parts with tools can be along with hemodialysis machine can be set along. 5,Rich knowledge engineer and sales can be serve for you.6,One year guarantee time,5years spare parts supplied at least.
Forms of Therapy for Hemodialysis Machine: 
  Haemodialysis, Isolated Ultrafiltration, Sequential Ultrafiltration, Hemoperfusion, etc.
Main Features:
  Intelligent Double Operation System 
  LCD touch screen with button interface 
  Emergency power 30Mins (Optional) 
  Blood Pump 
  Spare Pump (for standby and also can be used for hemoperfustion).
  Heparin Pump
  Hydraulic compartment(Balance chamber + UF pump).
  Operation, Alarm information Memory function.
  A/B Ceramic proportion pump , High precision, Corrosion-proof, Accuracy.
  
  Introduction 
  Size & Weight Size: 380mm×400mm×1380mm (L*W*H) 
  Area: 500*520 mm 
  Weight: 88KG
  Power Supply AC220V, 50Hz / 60Hz, 10A
  Input power: 1500W
  Back-up battery: 30 minutes (optional)
  Water input pressure: 0.15 MPa   ~0.6 MPa
  21.75 P.S.I. ~87 P.S.I.
  Water input temperature: 10℃~30℃
  Working environment: temperature 10ºC ~30ºC at relative humidity of no more than 70%.
  
  Dialysate
  Dialysate temperature: preset range 34.0℃~39.0℃
  Dialysate flux: 300~800 ml/min
  Dialysate concentration: 12.1 mS/cm ~16.0 ms/cm, ±0.1 ms/cm
  Dialysate mixing ratio: can set variety ratio.
  UF rate Flow range: 0 ml/h ~4000 ml/h
  Resolution ratio: 1ml
  Precision: ±30 ml/h
Extracorporeal Part
  Venous pressure: -180 mmHg ~+600 mmHg, ±10 mmHg
  Arterial pressure : -380 mmHg ~+400 mmHg, ±10 mmHg
  TMP  pressure : -180 mmHg ~+600 mmHg, ±20 mmHg 
  Blood pump flow range: 20 ml/min ~400 ml/min(diameter:Ф6 mm)
  Spare pump flow range: 30 ml/min ~600 ml/min(diameter:Ф8 mm)
  Resolution ratio: 1 ml
  Precision: error range ±10ml or 10% of reading
Heparin Pump
  Syringe size: 20, 30, 50 ml
  Flow range: 0 ml/h ~10 ml/h
  Resolution ratio: 0.1ml
  Precision: ±5%
Sanitize 
  1.Hot decalcification :
  Time: about 20 minutes
  Temperature: 30~60℃, 500ml/min.
  2.Chemical disinfection :
  Time: about 45 minutes
  Temperature: 30~40℃, 500ml/min.
  3. Heat disinfection :
  Time: about 60 minutes
  Temperature: >85℃, 300ml/min. 
  Storage Environment  Storage temperature should be between 5℃~40℃, at relative humidity of no more than 80%.
Monitoring System 
  Dialysate temperature: preset range 34.0℃~39.0℃, ±0.5℃
  Blood leak detection: Photochromic
  Alarm when erythrocyte specific volume is 0.32±0.02 or blood leak volume is equal or more than 1ml per liter of dialysate
  Bubble detection: ultrasonic
  Alarm when a single air bubble volume is more than 200µl at 200ml/min blood flow
  Conductivity: acoustic-optic, ±0.5%
All disposable accessories or other accessories such as Reprocessor,Dialyzer Rinse Machine,RO Water Machine,Dialysate Dosing Device,Dialysis Chair,Dialyzer,Hemoditoxifier,Anti-cross Infection Joint,Test Paper,Dialysis Catheter,Disposable A.V. Fistula Needel,Disinfectant,Dialysis Concentrate and Powder,Haemodialysis tubing package,etc.
Note:One year quality guarantee and 24 hours phone service for all the products supplied.

 
					