prev
  • Laboratory Automatically Microplate Processor Workstation
next
Prev | Next
SKU: HP-EIA1000
Laboratory Automatically Microplate Processor Workstation
Description
HP-EIA1000 Automatical Microplate Processor is fully automated two plates ELISA processing work station.It combine capable of sampling,incubation,washing and date processing in one machine.Easy operation and safety usage.It can be complete supply solution
χ
This item has added to inquiry basket

Please check your inquiry basket.

χ
This item was already in your inquiry basket

Please check your inquiry basket.

  • Product Detail
  • Send Inquiry

HP-EIA1000 Automatical Microplate Processor is fully automated two plates ELISA processing work station.It combine capable of sampling,incubation,washing and date processing in one machine.Easy operation and safety usage.It can be complete supply solution for your IVD ELISA assays.

Main Advantages:

2 microplates, 80 sample rack, 92 dilution rack.
8 special reagent racks, 1 general reagent racks.
Biological safety protection provides the high level of operator and environmental protection.
User-friendly Windows system; LIS system available
Hydrophobic membrane to prevent cross contamination sample module
Original sample tube and reagent bottles;Onboard large capacity and optional programmable dilution module.
Unique integrative washing and reading module system,High precision optics

Main Technical Parameters:

Model

HP-EIA1000

Sample Pipetting Unit

Sample Rack

 80 sample positions(tubes 12mm)

Primary Sample Tubes

12*75mm

Pipetting

10ul~1000ul, (100ul) CV<0.5%

Dispensing System

1 aspirating-dispensing probe with X-Y-Z    movement; dilution system with a syringe of 1000ul, continuous flow washing    station.

Reagent Pipetting Unit

Reagent Racks

8 special reagent racks,1 general reagent    racks

Removable Trays

Modular racks to hold calibrators,    controls and reagents for 8 test on line

Pipetting

10ul~1000ul,1ul stepping

Pipetting precision

(100ul) CV<0.5%

Time to Dispense

4 minutes adding 96 reagents

Incubation Unit

Incubators

2 independent units

Temperature Range

RT to 45 °C

Temperature Accuracy

±0.5°C

Temperature Fluctuation

±0.5°C

Washing Unit

Programmable

Yes(options & parameters)

Wash Containers

2 wash buffers at 2.0L, with    level-sensing

Washing System

Two 8 nozzle manifolds, one to dispense    and the other one to aspirate

Washing Residual

<2μl

Reading Unit

Reading System

8 independent photometric channels with    mono and dichromatic reading capability

Photometric Range

0.000 ~3.000 OD

Spectral Range

400-700nm

Optical Filters

2 precision filters (450,630nm); 2 more    on request (4 filters max, 400-700nm)

Software Unit

Software System

Windows 7/8/10 or above, LIS system    available

Function Application

Work list set up, patient reports and    result filing; calibration curve storage for each carried out test

Data Reduction

cut-off (qualitative)

Physical Specification

Package Size(W*D*H)

1050*800*1280mm

Gross Weight

98kgS

Power Supply

AC220V/50;110V/60Hz(Optional); 500W


Note:1,One year quality guarantee and 24 hours phone service for all the products supplied.2,Training Service Supplied!