prev
  • 2 pipetting probes Automatially ELISA Processor Workstation
next
Prev | Next
SKU: HP-EIA2000
2 pipetting probes Automatially ELISA Processor Workstation
Description
HP-EIA2000 Automatical Microplate Processor is fully automated two plates ELISA processing work station.It combine capable of sampling,incubation,washing and date processing in one machine.Easy operation and safety usage.It can be complete supply solution
χ
This item has added to inquiry basket

Please check your inquiry basket.

χ
This item was already in your inquiry basket

Please check your inquiry basket.

  • Product Detail
  • Send Inquiry

HP-EIA2000 Automatical Microplate Processor is fully automated two plates ELISA processing work station.It combine capable of sampling,incubation,washing and date processing in one machine.Easy operation and safety usage.It can be complete supply solution for your IVD ELISA assays.

Main Advantages:

 4 microplates,2 pipetting probes
 192 sample positions , 96 dilution positions,23 reagent   positions, 16 common reagent positions,48 QC positions.
 All sample & reagent positions can be editable.
 User-friendly Windows system; LIS system available
 Hydrophobic membrane to prevent cross contamination sample module
 Original sample tube and reagent bottles;Onboard large capacity and optional programmable dilution module.
 Unique integrative washing and reading module system,High precision optics
 
Main Technical Parameters:

Model

 HP-EIA2000

 

Sample pipetting unit

 Sample rack

 24 sample tubes/rack, at most    23 racks

 Primary sample tubes

 11.5-13mm

 Pipetting

 8ul~1000ul, (10ul) CV<5%

 Time to dispense

 12 minutes adding 96    samples

 Dispensing system

 1 aspirating-dispensing probe with    X-Y-Z movement;dilution system
      with a syringe of 1000ul,    continuous flow washing station.

Reagent pipetting unit

 Reagent racks

 2 reagent tanks(60ml)/ rack, at    most 23 racks

 QC racks

 24 positions/QC rack, at most    23 racks

 Pipetting 

 8ul~1000ul,1ul stepping

 pipetting precision

 (100ul )CV≤0.5%

 Time to dispense

 3 minutes(typical)^100ul of 96    reagents to plate from reagent bottle

Washing unit

 Programmable

 Yes(options & parameters)

 Wash containers

 3 wash buffers at 2.0L,1 wash    buffers at 5.0L,with level-sensing

 Waste containers

 25L with waste full sensor

 Washing system

 two 8 nozzle manifolds, one to    dispense and the other one to aspirate

 Washing residual

 <2μl

Reading unit

 Reading system

 8 independent photometric channels    with mono
      and bichromatic reading capability

 Photometric range

 0.000 ~3.000 OD

 Spectral range

 400-700nm

 Optical filters

 2 precision filters (450and 630nm);    2 more on request
      (4 filters max, 405 nm, 450 nm, 492    nm and 630 nm)

 Reading inaccuracy

 inaccuracy±1% from 0.000 to 1.500    OD, ±2% from 1.500 to 3.000 OD

Incubation unit

 Incubators

 4 independent units

 Programmable

 Yes

 Temperature range

 RT to 45 °C

 Temperature accuracy

 ±0.5℃

 Temperature staility

 ±0.5℃

Software unit

 Software system

 Windows 7/8/107 or above,    LIS system available

 Function application

 work list set up, patient reports    and result filing; calibration curve storage for each  carried out    test

 Data reduction

 cut-off (qualitative)

Others

 Power Supply

 AC 110/220V±10%,50/60Hz,450W

 External Size(W*D*H)

 1240*700*890mm

 PackageSize(W*D*H)

 1400*880*1450mm

 Gross Weight

 230kgs

Note:1,One year quality guarantee and 24 hours phone service for all the products supplied.2,Training Service Supplied!